Electron Beam vs. Gamma Radiation Sterilization

What is sterilization, and why is it essential for medical devices?

Sterilization keeps patients safe from toxins and microbial illnesses when therapies or devices are consumed or used. Sterilization is any process that removes, kills, or deactivates all microbial life. Under the strictest definition of sterility, an item or product is sterile when there is the complete absence of viable microorganisms (bacteria, yeasts, viruses, and molds). Sterility is defined by regulatory acceptance criteria based on calculated contamination probability. An acceptable level of contamination risk for most items is the probability of a single contaminated product out of a million manufactured products. However, sterility criteria may be more stringent or lax depending upon the intended use of the medical device or product. 

What are the top medical device sterilization methods?

There are seven primary methods for medical device sterilization. These methods are steam sterilization, radiation sterilization, dry heat sterilization, sterilization by filtration, gas sterilization (such as ethylene oxide sterilization), vapor sterilization, and liquid sterilization. Of these sterilization methods, gas sterilization, vapor sterilization, and liquid sterilization techniques are chemical sterilization processes. As you might expect, steam and dry heat sterilization are thermal sterilization processes.

What is radiation sterilization (sterilization by radiation)?

Sterilization by radiation is a non-thermal sterilization method that functions by destroying any microorganisms in a product with gamma radiation, beta particles (electron beam), x-ray, or ultraviolet (UV) light. Other than sterile filtration, sterilization by radiation is the only other sterilization method that doesn’t rely on elevated temperature to sterilize. Sterilization by radiation is an excellent alternative for products that cannot be sterilized with heat or chemicals.

What is electron beam (e-beam) sterilization?

Electron beam (e-beam) sterilization has been gaining popularity for medical devices since its release in the 1960s due to its safety, speed, and negligible emissions. Electron beam sterilization is a form of radiation sterilization that utilizes beta particles to inactivate microbes. Electron beam systems scan medical devices with focused electrons (beta particles) to sterilize them. E-beam delivers higher dosage rates with less penetration compared to gamma radiation.

What is gamma sterilization?

Gamma is a radiation sterilization method. Gamma rays are created by the self-disintegration of Cobalt-60 (60Co) or Cesium-137 (137Cs). Gamma rays are highly penetrating and can inactivate microbes with more depth than e-beam beta particles. The thickness and density of a material determine the level of gamma penetration.

Some benefits of electron beam sterilization are:

  • E-beam sterilization is an internationally accepted and FDA-approved process.
  • E-beam has a high dosing rate and sterility assurance level (SAL), allowing for immediate release (no batch-to-batch sterility testing is needed after sterile processing)
  • E-beam can penetrate a variety of materials, including foils.
  • E-beam processes allow for temperature control during irradiation.
  • A well-controlled dose range can be achieved with e-beam sterilization.
  • E-beam sterilization is cost-effective.
  • E-beam sterilization is a fast process (a minute in small lots), allowing near-immediate access to fully sterilized products.
  • The speed of e-beam dosing protects the product’s material properties, prevents polymer degradation, and causes no damage to sterile seals on product packaging.
  • E-beam has a minimal atmospheric effect and only releases a slight amount of ozone.
  • E-beam sterilization doesn’t require a localized radioactive source.

Some of the drawbacks to electron beam sterilization are:

  • E-beam sterilization institution construction is expensive.
  • There are few e-beam radiation sterilization centers available for bulk sterilization.
  • E-beam sterilization is less penetrative than radiation sterilization utilizing gamma.
  • A concern when sterilizing finished products or active pharmaceutical ingredients (APIs) with beta particles is the risk of radiolytic byproduct formation (e.g., *OH) that could cause damage to the raw material, API, or product packaging system.

Some benefits of gamma sterilization are:

  • Gamma sterilization is an advanced technological method.
  • Gamma sterilization is a cold method, and the increase in temperature is slight.
  • Gamma has a high SAL.
  • Control of the gamma sterilization is easy and can only be made by varying the applied dose.
  • Gamma radiation has great penetration and can easily provide terminal sterilization for products in their final packages.
  • Gamma radiation can be used for drug delivery systems such as microspheres, liposomes, or monoclonal antibodies.
  • Gamma radiation processes are easily validated. Time changes for gamma sterilization cycles only shift when the 60Co source decomposes at a constant speed. Otherwise, gamma radiation is constant and can be controlled by controlling the dwell time/speed of the conveyor in every position while it turns around the radiative source.
  • Products sterilized with gamma radiation can be immediately released and do not require batch-to-batch sterility testing.
  • Gamma radiation exposure can reduce endotoxins (depyrogenate products).
  • Gamma radiation forms no residue after the sterilization process.
  •  

Some of the drawbacks to gamma sterilization are:

  • The dose rate is lower than electron beams.
  • Gamma radiation has limited dose flexibility.
Abstracte background with blur bokeh light effect. Electron beam vs gamma radiation sterilization. What is gamma radiation sterilization. What is gamma sterilization. Benefits of gamma sterilization. Drawbacks of gamma sterilization. How is gamma sterilization performed. What items can be sterilized by electron beam. What items can be sterilized by gamma radiation

What items can be sterilized by electron beam?

Standard devices and materials sterilized with electron beam are plastics, heat-labile materials, glass, and powders. E-beam irradiation can also be used for tissue materials like aortas, bone, cardiovascular valves, and hydrogels. Radiation damages the nucleoproteins of microorganisms, and thus, electron beam sterilization is not recommended for biologics.

What items can be sterilized with gamma radiation?

Gamma radiation can easily be applied to many materials. However, gamma sterilization is incompatible with polyvinyl chloride (PVC), acetal, and polytetrafluoroethylene (PTFE).

Standard devices and materials sterilized with gamma radiation are plastics, heat-labile materials, glass, and powders. Gamma is generally used to sterilize disposable medical equipment, medical devices, and implants such as arteriovenous shunts, peritoneal dialysis sets, aortic valves, peripheral vascular prosthesis, dental implants, and artificial eyelids. Radiation damages the nucleoproteins of microorganisms, and thus, gamma sterilization is not recommended for most biologics. However, certain devices with microspheres, liposomes, or monoclonal antibodies may be sterilized with gamma.

How is electron beam sterilization performed?

Simply speaking, e-beam sterilization is performed by exposing a product to beta particles. Beta particles are not electromagnetic and less penetrative than ionizing gamma radiation. However, most e-beam sterilization requires an electron accelerator, a rare and specialized machine. In electron accelerator systems, electrons are accelerated to near the speed of light and at high electron concentrations. Electron absorption by the product undergoing sterilization destroys the DNA of any live microbes. The mechanism of microbial DNA destruction is called DNA chain cleavage. DNA chain cleavage occurs by changing the DNA’s chemical and molecular bonds.

E-beam sterilization cycle characteristics depend on accelerated energy and adsorbed dose. The absorbed dose is the amount of interaction between the e-beam and the product that will be sterilized. The adsorbed dose is often defined as the absorbed energy per unit mass (Jewels per kilogram). Absorbed dose and absorption depth depend on the acceleration energy. Further, the microbial survival fraction is inversely proportional to the absorbed dose. Thus, the higher the absorbed dose, the smaller the fraction of surviving microbes.

Electron beam radiation effectiveness is dependent on the radiation dosage and time exposure. A 12-D sterilization overkill approach is often used for electron beam sterilization. The 12-D stands for providing a radiation dose sufficient to produce a 12-log reduction in the D-value of the most resistant microbial spore. Note that D-value determination for radiation uses dosage rather than time. A typical D-value range for the most resistant bacterial spore (Bacillus pumulis) to radiation is 1.7 to 2.0 megarad (MRad). A 12-D radiation dosage of 25 mRad is common, as this dosage is greater than 12-fold the D value of B. pumulis. However, the total radiation dose for e-beam sterilized products with tissue materials is in the range of 2 Mrad.

When a product goes through an e-beam sterilization conveyor (see Figure 1 below), the product moves under the e-beam at a set speed to obtain the desired electron dosage for the sterilization process. The total radiation dosage is distributed throughout the conveyor system so that there is no change in dose over time. An e-beam sterilization process can be adjusted by altering conveyor speed, which impacts the applied dose of the beam current.

Figure 1 Schematic of Radiation Sterilization Conveyor
Figure 1. Schematic of Radiation Sterilization Conveyor

Factors that affect electron beam sterilization:

  1. D value of the biological indicator or bioburden level of the item undergoing sterilization
  2. Radiation strength (accelerated energy)
  3. Radiation dose rate (absorbed dose)
  4. Conveyor speed

How is gamma sterilization performed?

Gamma sterilization is very similar to electron beam sterilization. The difference is that gamma rays created by the self-disintegration of Cobalt-60 (60Co) or Cesium-137 (137Cs) are used within radiation conveyors instead of beta particles. Gamma radiation also generally takes a longer exposure time than electron beam radiation. However, factors that affect electron beam sterilization and the overall sterilization process through a radiation sterilization conveyor system are the same. Gamma dosage is verified utilizing dosimeters stationed at various points throughout the radiation sterilization conveyor.

What are the differences between gamma and electron beam sterilization?

E-beam sterilization is superior to gamma radiation in several ways, including a smaller degradation effect on product materials due to shorter radiation exposure times. Further, e-beam sterilization cycles can be controlled using accelerated energy and adsorbed dose, whereas gamma radiation is primarily based on absorbed dose. The radiation strength for gamma sterilization systems cannot be manually adjusted like e-beam sterilization systems. Gamma radiation dose strength is lower compared to e-beam dose strength. As a result, gamma radiation processing is slower (minutes to hours), whereas e-beam sterilization is on the scale of seconds to minutes. However, gamma radiation is far more penetrative and can process larger items than e-beam sterilization methods, which only process small products. Further, industrial e-beam sterilization is harder to find than industrial-level gamma sterilization offerings.

Summary

Overall, medical devices, products, and therapies must be sterile. Sterilization is any process that removes, kills, or deactivates microbes. Electron beam sterilization is a form of radiation sterilization that utilizes beta particles to inactivate microbes. Gamma is also a form of radiation sterilization. However, gamma uses the self-disintegration of Cobalt-60 (60Co) or Cesium-137 to inactivate microbes. Both gamma and e-beam sterilization methods are alternatives to chemical and steam sterilization processes. Gamma sterilization systems can penetrate further into products and sterilize larger items. E-beam sterilization is less penetrative but a faster sterilization method with smaller degradation effects on product materials. All in all, ensure you choose a contract testing organization that can provide appropriate sterility testing for your product needs.

Ethide Labs is a contract testing organization that specializes in Sterilization Validations and Sterility Testing. Ethide Labs also offers Bioburden Testing, Microbiology Testing, Bacterial Endotoxin Testing, Ethylene Oxide Residual Testing, Cytotoxicity Testing, Environmental MonitoringPackage Integrity Testing services for medical device companies and allied industries. Ethide is an ISO 13485 certified facility. 

References

Michael J. Akers. Sterile Drug Products Formulation, Packaging, Manufacture, and Quality. Drugs and the Pharmaceutical Sciences. Informa Healthcare. 2010.

Mine Silindir and A. Yekta Ozer. Sterilization Methods and the Comparison of E-Beam Sterilization with Gamma Radiation Sterilization. FABAD J. Pharm. Sci., 34, 43-53. 2009.

United States Pharmacopeial Convention. <1211> Sterility Assurance. Rockville, MD, USA. 2021. (USPC <1211>).

United States Pharmacopeial Convention. <1229> Sterilization of Compendial Articles. Rockville, MD, USA. 2021. (USPC <1229>).

United States Pharmacopeial Convention. <1229.10> Radiation Sterilization. Rockville, MD, USA. 2021. (USPC <1229.10>).

Share this in your social networks